Evaluation of medical textiles

Textiles in operation theatres:

Requirements and conformity with the European Medical Devices Directive

As a medical device, textiles used in operation theatres have to be conform with the requirements of the European Medical Devices Directive 2007/47/EC (modified 93/42/EEC).

This conformity depends on the type of medical device and more in particular on the class to which it belongs. There are indeed four different classes of medical devices, the most severe one being class 4 for those medical devices including the highest risk of contamination.

Textiles used in operation theatres belong to class 1 (low risk) of the medical devices and are conform if they comply with the different parts of standard EN 13795. This standard has been written in the spirit of the European directive and aims at preventing the transfer of infections between the medical staff and the patient during surgical operations and other invasive interventions. At the same time, it is a fight against nosocomial (hospital related) infections.

Moreover, the standard wants to guarantee the same safety and performance levels for all products, disposable or reusable articles alike, and this during the entire product life.

EN 13795 deals with (woven or nonwoven) surgical drapes, gowns and clean air suits, used as medical devices, hospital staff and equipment. The standard defines the requirements and corresponding test methods. The standard was modified for the last time in 2009 and includes three parts: EN 13795-1:2002+A1 :2009, EN 13795-2 :2004+A1:2009, EN 13795-3:2006+A1:2009. A new version uniting these three parts is expected in April 2011.

Part 3 sets the minimum values for performances in compliance with the characteristics evaluated in part 1 according to the test methods described in part 2.

These minimum values account for the circumstances of the surgical operation. Therefore they distinguish between products for standard performances and products for higher performances on the one hand and between the critical and less critical zones of the product, on the other hand.

  • The critical zone of a product has a greater chance to be involved in the transfer of infection carriers on the operation site or on the invasive zone or vice versa. For example: the front piece and the sleeves of surgical gowns are in the immediate proximity of the operation site.
  • A product for standard performances or high performances is defined in function of the exposure to biological or other fluids, to mechanical pressure or to the duration of surgical operation

Which properties are evaluated according to EN 13795?

Part 2 of EN 13795 describes the following tests:

  • Tensile strength - dry and wet - EN 29073-3:1992
  • Burst strength - dry and wet - EN 13938-1
  • Resistance to fluid penetration - EN 20811
  • Cleanliness – microbial - EN 1174 (replaced by EN ISO 11737)
  • Cleanliness – particulate matter - ISO 9073-10
  • Linting - ISO 9073-10
  • Resistance to microbial penetration - dry - ISO 22612
  • Resistance to microbial penetration - wet - ISO 22610

ISO22610.png
ISO 22610: Resistance to microbial penetration - wet

ISO22612.png
ISO 22612: Resistance to microbial penetration - dry

ISO9073-10.png
ISO 9073-10 - linting

Protection of the patient versus protection of the medical staff

Standard EN13795 is aimed at the protection of the patient; this determines the direction in which the sample is brought into contact with the contaminating agent during the test.

If the manufacturer also claims the protection of the medical staff the surgical gown will no longer be considered as a medical device but as a PPE (personal protection equipment).

In this case, the product has to comply with the corresponding directive 89/686/EEC (protective clothing) and standard EN14126: Protective clothing. Performance requirements and tests methods for protective clothing against infective agents

Which additional properties are interesting?

Standard EN13795 mentions some interesting additional properties such as "liquid control" or adhesion properties.
Centexbel offers the manufacturer the following test methods to measure the comfort properties of his/her products:

  • Skin model – ISO 11092 - EN31092
  • Thermal manikin - EN ISO 15831 – ASTM F2370
  • Air permeability - ISO 9237
  • WP resistance to water - ISO 811

Besides tests that have to prove the conformity of the product, the European directive introduces the notion of risk analysis by proposing the following tests:

  • non-toxicity of the medical device – ISO 10993
  • measurement of the electrical risk– ISO 2878 (BS2050) and EN 1149
  • Protection against laser beams
    • ISO 11810-1 : primary ignition and penetration
    • ISO 11810-2: secondary ignition

Centexbel carries out all above-mentioned tests