Protective gloves against dangerous chemicals and micro-organisms according to ISO 374-2:2019
Part 2: Determination of resistance to penetration
Testing lab
Chemical lab
Legislation
PPE Regulation (EU) 2016/425
Testing standard
EN ISO 374-2:2019
Principle
Assessment whether a protective glove acts as an effective barrier against liquid and/or air penetration. Penetration occurs when chemicals, liquids, or microorganisms pass through porous materials, seams, pinholes, or other imperfections.
Air Leak Test (Inflation Test)
Purpose:
- Detect air leakage through pinholes or seams.
Method:
- The glove is filled with air and submerged in water.
- Pressure is applied (typically 1–2 kPa).
- The glove is observed for any bubbles, which would indicate a leak.
Duration:
- Usually 30 seconds to 2 minutes.
Pass Criteria:
- No bubbles = pass (no leakage).
Water Leak Test
Purpose:
- Detect water leakage through defects in glove materials.
Method:
- The glove is filled with water and suspended vertically.
- It is observed for a fixed period (typically 2 minutes).
- Any droplet formation on the outer surface is considered a failure.
Volume:
- Usually filled with 1 liter of water (or to cuff level).
Pass Criteria:
- No droplets = pass.
Pass/Fail Criteria and Classification
Gloves are tested in accordance with ISO 2859-1 sampling procedures (based on statistical AQL – Acceptable Quality Level).
- Typical AQL Levels:
- Level 1: AQL 4.0 (low protection)
- Level 2: AQL 1.5 (medium protection)
- Level 3: AQL 0.65 (high protection)
To comply with EN ISO 374-2, no failures should exceed the AQL limit for the chosen performance level.